A patient with carbon monoxide poisoning has a dangerously low arterial oxygen content despite a pulse oximeter reading that appears falsely reassuring (near-normal SpO2). This occurs because standard two-wavelength pulse oximetry:
  • Cannot distinguish carboxyhemoglobin from oxyhemoglobin, reading both as if oxygen-saturated✔️
  • Directly measures and correctly subtracts carboxyhemoglobin from the saturation value
  • Is unaffected by hemoglobin binding characteristics of any kind
  • Accurately measures dissolved oxygen but not hemoglobin-bound oxygen
Correct Answer Explanation
Standard pulse oximeters use only two wavelengths and cannot differentiate carboxyhemoglobin from oxyhemoglobin, so they report a falsely near-normal saturation even though carbon monoxide has displaced oxygen from most hemoglobin binding sites and also shifts the curve left — true hypoxia is masked, and diagnosis requires co-oximetry.